Antong fully automatic tunnel blood pressure monitor
AMPall BP868F
1) Detection valve: Connect a mercury column blood pressure meter or precision pressure meter to check the accuracy of the measurement value.
2) RS232C transmission line interface: The transmission line interface connected to the computer, the measurement data is transmitted to the computer for management of personal information.
3) Voice prompt switch: Turn on/off the voice system and adjust the volume.
4) Print switch:
ON: Automatically print the measured value after measuring blood pressure.
OFF: Turn off the printer. If you need to print data, please press the print button.
5) Ground wire interface: This instrument has an insulated structure, which does not pose a safety hazard to the user. In order to prevent leakage accidents caused by the use environment, a ground wire interface is also provided.
6) Power interface: Connect the power cord
7) Power switch: ON/OFF of the device
Thermal paper: After taking the blood pressure, the thermal printing paper must be used when printing the measured values. Length: 25M Width: 57.5mm
Ministry of Health Medical Equipment Transmission No. 028772
model | BP868F |
Measurement method | Oscillometric method |
Exhaust method | Microcomputer automatic control |
pressure measurement | semiconductor sensor |
Display method | LCD或FND (7-Segment) |
Measuring range | Blood pressure: 30~300mmHg, pulse: 30~200 beats/min |
smallest unit of measurement | 1mmHg |
Precision | Blood pressure: 2mmHg, pulse: 3% |
Cuff drive method | Microcomputer control driver |
Pressurization time | about 10 seconds |
Measurement time | about 30 seconds |
Printing method | High-speed thermal printing |
power supply | AC 110V |
frequency | 50 / 6 0 Hz |
Communication function | Can be connected to computer (optional RS232C transmission line) |
emergency stop function | Microprocessor driven emergency stop |
Voice prompt function | Provide voice guidance during blood pressure measurement/provide blood pressure value and fault voice prompts after measurement |
power | 30VA |
Operating environment requirements | Temperature +10. C ~ +4 0. C, humidity below 95% |
Storage environment requirements | Temperature -5. C ~ +5 0. C, humidity below 95% |
Volume size | Width 46cm * Depth 31cm * Height 29.6cm |
weight | Approximately 9Kg |
Accessories | Power cord, printing paper, arm tray, user manual, operating guide (optional) |
Other options | Customer management program, transmission line, automatic cutting and printing, rear display panel |